WASHER DETERGENT – NEODISHER Z NEUTRALISER

5L
GKE-WE420233

Acidic neutralising agent and detergent for the automated reprocessing of instruments and laboratory glassware  

 Fields of Application 

  • Neutralisation of alkaline residues during automated reprocessing of surgical instruments, anaesthesia utensils, containers, OR shoes and baby bottles as well as other medical utensils 
  • Neutralisation or acidic pre-cleaning for the automated cleaning of laboratory glassware in industrial, water and medical laboratories  

Performance Spectrum 

  • Neutralises alkaline residues from the main cleaning step during automated reprocessing  
  • Reliably removes acid-soluble residues during pre-cleaning  
  • Excellent material compatibility; therefore suitable for anaesthesia utensils, surgical instruments and other sensitive materials  
  • Not suitable for acidic pre-cleaning of objects and parts made of anodised aluminium as well as chromed and nickelplated parts  

 

Special Properties  

  • Based on organic acids  
  • Free of phosphates, phosphoric acid, nitrogen compounds and surfactants, therefore no disturbing influence on analytical test methods  
  • Contains less than 10 ppm P2O5 in the concentrate 

 

Application and dosage 

Neodisher Z is used in washer disinfectors for neutralisation and/or acidic pre-cleaning. The dosing amount depends among other things on the application.  

When using neodisher Z the following parameters are recommended 

For neutralisation  1 – 2 ml/l 
For acidic pre-cleaning/cleaning  2 – 5 ml/l, 40 – 60 °C 

Suitable dosing devices are to be used.  

For avoiding water stains, the use of deionised water in the final rinse is recommended. At the same time this protects anodised aluminium.  

When using neodisher Z for pre-cleaning the items to be washed, the washer resp. washer disinfector and the waste water pipes must be acid-compatible.  

Eternit (fibre cement) and cast iron pipes are unsuitable and require a prior neutralisation of the effluent solution.  

 

General notes on Application 

  • For professional use only.  
  • Do not mix with other products.  
  • The neodisher Z solution must be rinsed off completely (preferably with deionised water).  
  • Rinse out dosing system including suction hose with water before changing product.  
  • Reprocessing should comply with all ordinances pursuant to the regulations on medical devices and should be performed with appropriate validated processes. 
  • Observe the instructions given by the manufacturers of the washer disinfector. 