MISTRAL MA1200 AIR WARMING BLOWER

WITH 2 BLANKET SETS
210162-K1

Experience the benefits of the Mistral MA1200 patient warming system, including:

  • Improved patient outcomes
  • Compact, lightweight and user friendly design
  • Easy to clean unit, made of shock-absorbent material
  • Optimised heat transfer with a powerful fan
  • Optimal elimination of air contaminants by inbuilt the HEPA filter
  • Equipped with a versatile mounting bracket, the slim Mistral-Air warming unit offers caregivers a diversity to securely position the warming unit across all care areas.

 

Comes with 2 sets of washable blankets (2 Small, 2 Medium and 2 Large)

General specifications
Article number MA1200-EU
Rated voltage 220-240 V~
Frequency 50/60 Hz
Device sound pressure 48 dBA
Average current 3.2 A
Peak current 9.0 A
Peak power 1000 VA
Average power at 43°C and tamb 22 +/- 1.5°C 750 VA / 600 W
Fuses 6.3AHF/250V~
Air flow rate at nominal 86 – 101 [Nm³/h], 50 – 59
voltages and ambient [CFM] (MA1200-QC-EU: 60 –
temperature (1.8 m hose ) 101 [Nm³/h], 35 – 59 [CFM]),

(depending on orientation of the hose, supply voltage, type and drape of the blanket)

GMDN code  36954 (circulating-air whole-body heating system control unit)
Dimensions   16 cm x 35 cm x 40 cm (l x w x h)
Weight     5.2 kg
Hose length           1.8 m (3 m hose optional)
Power supply cord length   4.0 m
Filtration   HEPA H13 class filter, conform EN 1822-1:2009
Classification93/42/EEC     Class IIb
Classification IEC 60601-1   class I, Body Floating (BF)
Overvoltage category according to IEC 60664-1

 

Category II
Classification IEC 60529    IP23
Setpoint temperature      Ambient air, 32°C, 38°C, 43°C
Accuracy of temperature at the end of the hose ± 2.5 °C (under all validated operating environmental conditions)
Setpoint reached after      Under 30 seconds
Low temperature limit      10°C
Maximum average contact surface temperature 45.5°C (Compliant with IEC 80601-2-35)

 

High temperature safety limit  < 56°C (Compliant with IEC 80601-2-35)

 

Auditory alarm signal sound pressure 80 dBA
Applicable technical standards  IEC 60601-1:2005+A1:2012, IEC 80601-2-35:2016
Expected lifetime device     7 years
Expected lifetime hose  1 year
The essential performance of the Mistral-Air Warming System is: when supplying air to the patient “the Maximum CONTACT SURFACE TEMPERATURE” must be below the safe temperature limits according to IEC 80601-2-35: 2009 + C1:2012 +C2:2015, Clause 201.11.1.2.1.102.

 

Validated operating environmental specifications
Ambient temperature 15°C to 30°C
Relative humidity 30% to 75%
Validated transport and storage specifications
Ambient temperature -20°C to 70°C
Relative humidity 10% to 90% (non-condensing)
Atmospheric pressure 50 kPa to 106 kPa
2 Year Limited Warranty

Request a quote or more information​ about this product