CHAM-MATE Cleaning Procedures
• Basically CHAM-MATE is functioned in water
• 1. For Class-B vacuum functioned autoclaves, or autoclaves with independent steam generator, the CHAM-MATE must be mixed with water first. (1 x 30 g bag mixed into 2 Litres of water) and the, pouring the mixed CHAM-MATE into the chamber, and start the working cycle
• 1-1. For Class-S no vacuum autoclaves (with auto-add water function), put the CHAM-MATE powder into the chamber directly, and start the working cycle
• 1-2. For Class-N (manual add water function), put the CHAMMATE powder into the chamber directly, and then pouring water into the chamber and start the working cycle
• PLEASE NOTE: Running one sterilization cycle, (without the dry function)
• 2. Drain off all the balance water from the chamber and water reservoir
• 3. Open the door, wipe the chamber and the door with a wet, non-lint cloth
PS: Use one more bag of CHAM-MATE and recycle again as above procedures, if there are dregs in the chamber still
• NOTE: “For chamber and piping system clean only”
• CHAM-MATE quantity to be used:
• Chamber size under 30 litre – 1 bag CHAM-MATE
• Chamber size 30-100 litre – 2 bags CHAM-MATE
• Chamber size over 100 litre – 1 bag more per 100 litre more